Food and Drug Administration is reminding patients who use the popular 600-series MiniMed insulin pumps made by Medtronic that the devices contain an issue that could cause their buttons. In this video Kristen starts up her new Medtronic Minimed 670g for the first time. She will show you the start up menus as well as setting a basal rate with.
Table S2 | Years in which the reported trouble occurred.
Table S3 | Details of the batteries in the reported trouble cases.
Table S4 | Types of trouble related to the batteries and the power units.
Table S5 | Types of error message.
Table S6 | Actions required to resolve the trouble.
GUID: 3CF02DCA-8784-4ABD-9D03-088C6BE40F59
Abstract
Aims/Introduction
We investigated the safety of the batteries and power units used in insulin pumps in Japan.
Materials and Methods
A self‐administered questionnaire was sent to the 201 members of the Association for Innovative Diabetes Treatment in Japan.
Results
A total of 56 members responded, and among the 1,499 active devices, 66 had episodes of trouble related to the batteries and power units. The ratio of reported troubles to the number of insulin pumps was significantly higher in insulin pumps with a continuous glucose monitoring sensor compared with insulin pumps without a continuous glucose monitoring sensor (odds ratio 2.82, P < 0.05). The cause and the consequences varied. The brands of the batteries varied; alkaline batteries purchased at drug stores and other shops accounted for 19.7%. Termination of battery life within 72 h of use was reported most frequently (50.0%), suspension of the insulin pump (21.2%) and leakage of the battery fluid (4.5%) followed. A total of 53.2% of the reported insulin pumps needed to be replaced, and 37.1% of them recovered after replacement of the battery.
Conclusions
As trouble related to the batteries and power units of insulin pumps was frequent, practical guidance should be provided to respective patients regarding the use of reliable batteries, and to be well prepared for unexpected insulin pump failure.
Keywords: Battery, Continuous subcutaneous insulin infusion, Insulin pump
Introduction
Insulin pump therapy, or continuous subcutaneous insulin infusion, consists of an infusion set with an insulin reservoir, a tube and a cannula, pumping mechanism, computer chip, screen, casing, and battery. The prevalence of insulin pump therapy for patients with type 1 diabetes mellitus is increasing in many countries. Rapid‐acting insulin analog is used in the insulin pump, and it is infused subcutaneously to mimic physiological insulin secretion1. Pharmacokinetic studies showed that the rapid‐acting insulin analog acts within 15 min, peaks in 1 h and disappears within 4 h after subcutaneous injection2. Insulin infusion failure is a critical problem in insulin pump therapy because of a small amount of subcutaneous residual insulin3, 4. Once the insulin infusion is accidentally stopped, patients start to develop diabetic ketoacidosis within 5–6 h5. Therefore, it is very important that insulin infusion is maintained during insulin pump therapy. We sometimes experience diabetic ketoacidosis caused by insulin pump malfunctions, such as the obstruction of the infusion set or the dislocation of the cannula from the skin, and, as the majority of type 1 diabetes mellitus patients require insulin to maintain their lives, the loss of function of insulin pumps can become a fatal problem6. However, the incidence and the causes of battery and power unit failure during insulin pump therapy have been rarely published to date. Batteries sold at drug stores and other shops are cheap, and sometimes the manufacturers are not identifiable. Therefore, the Association for Innovative Diabetes Treatment in Japan (ASINDTJ) decided to investigate the prevalence of trouble regarding batteries and power units used in insulin pumps.
Methods
Study design and participants
A multicenter, cross‐sectional survey regarding the safety of batteries and power units used in insulin pumps was carried out among ASINDTJ members. The questionnaire used for the survey was designed by the authors, and underwent content and construct validation by the organizers of the association. A self‐administered questionnaire was mailed to all the 201 members (physicians 85.6%, registered nurses 3.5%, pharmacists 3.0%, laboratory technicians 3.0%, registered dieticians 2.0%, other occupations 3.0%) across Japan in 2016. The following variables were assessed: type of insulin pump, number of battery and power unit problems, combination of continuous glucose monitoring (CGM; yes/no), year of the problem, type of battery, cause of the problem and actions required to resolve the troubles.
Statistical analysis
All statistical analysis was carried out using R (R Foundation for Statistical Computing, Vienna, Austria)7. We reported data for categorical variables, such as percentages, odds ratios and 95% confidence intervals (CI). We compared the characteristics of insulin pumps with or without a CGM sensor using Fisher's exact test. P‐values of <0.05 were considered significant.
Ethics
As the present study was strictly limited to information about insulin pumps and contains no information about the patients wearing them, ASINDTJ decided that this study did not need to be approved by an ethics committee.
Results
A total of 56 members of the ASINDTJ responded to the questionnaire. As of 1 September 2016, MiniMed 620G, Paradigm 722 and Paradigm 712 (Medtronic MiniMed, Inc., Northridge, CA, USA) were mainly used in Japan, and a small number of TOP‐8100 and TOP‐8200 (TOP Corporation, Tokyo, Japan) were used (Table S1). Just 14.3% of the respondents supplied the batteries for insulin pumps to patients. Among them, 12.5% purchased the batteries at a medical institution, and 87.5% were provided batteries by the sales representatives involved in the leasing contract of the insulin pumps. Regarding the supplier of the batteries, 62.5% wished that the batteries be included in the leasing contract of the insulin pumps. Regarding reasons, 80% expected the supply of appropriate batteries, 65.7% thought pump users would not have to purchase the batteries themselves and 60% expected the possibility of using poor quality batteries would be reduced (multiple answers were accepted).
A total of 24 ASINDTJ members reported troubles occurring in the batteries and power units, and the reports on MiniMed 620G with CGM sensor were most common (Table (Table1).1). There were 44 cases out of 639 insulin pumps with a CGM sensor ([1] in Table Table1),1), and 22 cases out of 860 insulin pumps without a CGM sensor ([2]–[7] in Table Table1).1). The ratio of reports regarding problems related to the batteries and the power units to the number of insulin pumps was significantly higher in insulin pumps with a CGM sensor compared with insulin pumps without a CGM sensor (odds ratio 2.82, 95% CI 1.67–4.75, P < 0.05). The reported problems were mainly in 2015 and 2016, and many of the problems occurred in the MiniMed 620G with a CGM sensor (Table S2).
Table 1
Model | No. reports | No. active devices | Rate (%) |
---|---|---|---|
(1) MiniMed 620G with CGM sensor | 44 | 639 | 6.9 |
(2) MiniMed 620G without CGM sensor | 4 | 207 | 1.9 |
(3) Paradigm 722 | 14 | 568 | 2.5 |
(4) Paradigm 712 | 3 | 64 | 4.7 |
(5) TOP‐8200 | 0 | 4 | 0.0 |
(6) TOP‐8100 | 0 | 2 | 0.0 |
(7) Others | 1 | 15 | 6.7 |
Total | 66 | 1,499 | 4.4 |
(1) Insulin pumps with continuous glucose monitoring (CGM) sensor. (2)–(7) Insulin pumps without CGM sensor.
The brands of the batteries varied; and alkaline batteries purchased at drug stores, hundred‐yen stores, volume‐sales electronics retailers and DIY stores accounted for 19.7% of the report (Tables (Tables22 and S3). Regarding the types of trouble related to the batteries and the power units, termination of the battery life within 72 h of use was reported most frequently (50.0%), suspension of the insulin pump (21.2%) and leakage of the battery fluid (4.5%) followed (Tables (Tables33 and S4). An error message generated by the MiniMed 620G with a CGM sensor, ‘Error 25’ (power error detected), was reported most frequently (Tables (Tables44 and S5). Regarding the actions required to resolve the trouble, 53.2% of the reported insulin pumps needed to be replaced, 37.1% of them recovered after replacement of the battery and 9.7% of them recovered by following the manufacturer's instructions, including telephone support (Table S6).
Table 2
Details of the batteries in the reported trouble cases
Type of battery | Brand and product name of the battery | n | Rate (%) |
---|---|---|---|
(1) Lithium battery | Energizer | 2 | 3.0 |
Panasonic | 7 | 10.6 | |
(2) Alkali battery | Panasonic Evolta | 13 | 19.7 |
Panasonic except Evolta | 7 | 10.6 | |
Panasonic (unknown product name) | 4 | 6.1 | |
Fujitsu Premium | 1 | 1.5 | |
Maxell Voltage | 1 | 1.5 | |
Toshiba Alkali1 | 3 | 4.5 | |
Batteries purchased at drug store, hundred‐yen store, volume‐sales electronics retailer and DIY store | 13 | 19.7 | |
Unknown | 2 | 3.0 | |
(3) Rechargeable battery | Panasonic Rechargeable Evolta | 1 | 1.5 |
Unknown | 1 | 1.5 | |
(4) Unclassifiable battery | Unknown | 13 | 19.7 |
Total | 68 | 100 |
Table 3
Types of trouble related to the batteries and the power units
n | Rate (%) | |
---|---|---|
(1) Termination of battery life within 72 h of use | 33 | 50.0 |
(2) Suspension of the insulin pump | 14 | 21.2 |
(3) Leakage of the battery fluid | 3 | 4.5 |
(4) Usage of inappropriate batteries | 1 | 1.5 |
(5) Others | 15 | 22.7 |
Total | 66 | 100 |
Table 4
Types of error message (MiniMed 620G with continuous glucose monitoring sensor)
n | Rate (%) | |
---|---|---|
(1) Power error detected (Error 25) | 17 | 38.6 |
(2) Insert battery (Delivery stopped) | 2 | 4.5 |
(3) Low battery (Replace battery soon) | 15 | 34.1 |
(4) Replace battery now (Delivery stopped) | 4 | 9.1 |
(5) Replace battery (Battery life less than 30 min) | 2 | 4.5 |
(6) Battery not compatible | 0 | 0.0 |
(7) Battery failed (Insert a new battery) | 1 | 2.3 |
(8) Pump error (Delivery stopped) | 1 | 2.3 |
(9) Others | 4 | 9.1 |
Total | 46 | 100 |
![Medtronic minimed 620g Medtronic minimed 620g](/uploads/1/1/8/0/118013302/315461852.gif)
Discussion
In the present study, the rate of trouble related to the batteries and power units used in insulin pumps was approximately 5% among all active devices. The ratio of reports related to the batteries and power units of the insulin pumps to their total number was significantly higher in insulin pumps with a CGM sensor (sensor‐augmented pumps) compared with insulin pumps without a CGM sensor.
The reported troubles occurred mainly in 2015 and 2016, and this might be related to the fact that sensor‐augmented pumps became commercially available in Japan from February 2015.
As this survey does not include information regarding the numbers, periods, manufacturers and brands of the batteries used in pumps without the batteries and the power unit problem, the rate of occurrence is unclear. However, considering the generally accepted idea that the batteries used in insulin pumps should be reliable, it might be better to recommend that insulin pump users choose batteries from well‐known manufacturers rather than those of unidentified manufacturers.
It is important to note that approximately 20% of the report resulted in the suspension of the insulin pumps, which could potentially have caused severe adverse events including hyperglycemia, diabetic ketoacidosis or death in absolutely insulin‐dependent patients if appropriate actions, such as changing the batteries or injecting insulin using alternative devices, had not been taken within an adequate time to resolve the trouble. It is also important to note that approximately 5% of the reports resulted in the leakage of battery fluid inside the insulin pumps, which could potentially lead to a complete loss of function of the devices including the alarm function.
Batteries are not the only cause of power unit‐related problems. For example, approximately 40% of the error messages generated by MiniMed 620G with a CGM sensor were ‘Error 25.’ This ‘Error 25’ is caused by the interference of the software operating the pump, and indicates drainage of the internal rechargeable battery due to the disturbance in its charging process. Some of the early termination of battery life might be related to the software that diagnoses the decline of battery power. As MiniMed 620G pumps were designed and built in the USA, it is not clear which battery available in Japan is really suited to use with the device. Medtronic started to sell alkaline batteries manufactured by Energizer (St. Louis, MO, USA) as the recommended battery to use with its pumps from January 2017 in Japan. Further evaluation is required to specify other recommendable batteries from different manufacturers.
The reason why almost half of the insulin pumps with battery and power unit problems needed to be replaced in the present study is unclear. Medtronic disclosed that it replaces insulin pumps with multiple episodes of ‘Error 25.’ Medtronic has updated the software version of the MiniMed 620G from 2.6B to 2.9B to reduce ‘Error 25’; however, the effect is unknown at this moment.
From the present study, we would like to propose that battery‐ and the power unit‐related problems are frequent events. It is worth noting the special background of insulin pump prescription in Japan. Unlike in the USA, insulin pumps are leased to patients from medical institutions due to the monthly health insurance reimbursement system. The majority of the insulin pumps are subleased from medical institutions, to which insulin pumps are leased by the representative sales companies. Some of the insulin pumps are purchased by the institutions and leased to the patients, but this is a minority. When governmental regulations regarding the health insurance reimbursement of insulin pump therapy are strictly interpreted, medical institutions need to supply all the consumables to the patients. This means not only the infusion sets, but also batteries need to be provided to the patients; however, the reality is that only a limited number of medical institutions are executing this, as shown in the present study.
There are several previous studies that surveyed insulin pump failure; however, detailed information related to batteries and power units is limited. A prospective study carried out in France reported that there were 26 cases of ‘defect in reservoir or battery compartment’ among 42 mechanical defects, accounting for 18% of the 232 entire insulin pump failures including complete pump failure, alarm set off and minor defect8. A cross‐sectional study using a self‐report questionnaire completed by patients in the UK reported ‘battery compartment problem’ accounted for 11% of the insulin pump malfunction9. However, in these studies, it was unclear how much of the battery compartment problem was related to the batteries themselves rather than the problem of the insulin pump enclosure. A prospective study of pediatric and adolescent patients in Australia and New Zealand surveying various adverse events including pump malfunctions, set/site related, severe hypoglycemia and cutaneous problems reported adverse events up to 40 out of 100 person‐years, of which ‘battery issue’ accounted for 6%; that is, frequency of 2.4 out of 100 person‐years10. Another prospective study carried out in France reported that the incidence rate of insulin pump malfunctions was 33 out of 100 pump‐years, and mechanical defect accounted for 35% of the reported malfunctions; however, the proportion of the trouble related to batteries and power units was not described11. In the present study, 53.2% of the insulin pumps with problems in the batteries or power units needed to be replaced. A retrospective study carried out in Italy reported that 0.165 insulin pump replacements per patient‐year were experienced in children and adolescents with type 1 diabetes, and Animas VIBE (Animas Corporation, West Chester, PA, USA) and MiniMed VEO (Medtronic MiniMed, Inc., Northridge, CA, USA), both with CGM connectivity, were the most replaced in this survey, but the difference between sensor‐augmented pumps and insulin pumps without a CGM sensor did not reach statistical significance12.
A limitation of the present study was its cross‐sectional observational design, lacking information about the frequency of troubles and patient background, together with a low response rate (27.9%). There is the possibility of missed reports that can affect the number of incidents, because the self‐administered questionnaire was not sent to the patients directly, and the actual number of incidents might be much larger than that described in this report. A prospective study regarding the safety of the batteries and power units used in insulin pumps would be valuable to further clarify the frequency of such problems.
As the troubles related to the batteries and power units of insulin pumps were frequent, practical guidance should be provided to respective patients to use reliable batteries, and to be well prepared for unexpected insulin pump failure. As some of the insulin pumps recovered after following the manufacturer's instructions, reminding patients to keep the telephone number of the manufacturer's telephone support at hand might be helpful. If the electricity supply to the insulin pump is completely lost, the alarm function could no longer be activated. Executing self‐monitoring of blood glucose four times a day or more, a standard procedure required for safe continuous subcutaneous insulin infusion, might help in finding insulin pump failure in such situations. Once insulin pump failure occurs, injecting insulin with insulin pens is important to prevent hyperglycemia and diabetic ketoacidosis. So far, the safety of the battery and power units used in insulin pumps has not been widely discussed. However, we believe that this is a very important issue related to the safety and the effectiveness of continuous subcutaneous insulin infusion.
Disclosure
The author(s) declare the following potential conflicts of interest to the research, authorship and/or publication of this article. Takashi Murata discloses the following relationships: lecture fees from Novo Nordisk, Eli Lilly, Sanofi, Arkray, MSD and Sumitomo Dainippon; and grants paid to the institution as study physician by Novo Nordisk, Sanofi, Kissei, Ono and, Astellas. Naoki Sakane discloses the following relationships: lecture fees from Novo Nordisk, Eli Lilly and Sanofi. Akio Kuroda discloses the following relationships: lecture fees from Eli Lilly and Sanofi. Yushi Hirota discloses the following relationships: lecture fees from Novo Nordisk, Eli Lilly, Sanofi, Medtronic and Abbott. Munehide Matsuhisa discloses the following relationships: lecture fees from Mitsubishi Tanabe, Eli Lilly, Astellas, Novo Nordisk, Sanofi and Novartis; and research grants from Daiichi Sankyo, Astellas, Mitsubishi Tanabe, Boehringer Ingelheim and Tokushima Data Service. Mitsuyoshi Namba discloses the following relationships: lecture fees from Sanwakagakukenkyusho, Sanofi, Novo Nordisk, Astra Zeneca, Novartis, Eli Lilly, Mitsubishi Tanabe and Kowa; and research grants from Arkray, Astellas, Novo Nordisk, Sanwakagakukenkyusho, Kyowa Hakko Kirin, Kowa, Ono, Daiichi Sankyo, Teijin, Mitsubishi Tanabe, Boehringer Ingelheim, Takeda, Sanofi, Sumitomo Dainippon, Astra Zeneca and Pfizer. Tetsuro Kobayashi discloses the following relationships: lecture fees from Sanofi. Shinsuke Nirengi declares no conflict of interest.
Supporting information
Table S1 | Number of insulin pumps and their models.
Table S2 | Years in which the reported trouble occurred.
Table S3 | Details of the batteries in the reported trouble cases.
Table S4 | Types of trouble related to the batteries and the power units.
Table S5 | Types of error message.
Table S6 | Actions required to resolve the trouble.
Acknowledgments
This study was supported by ASINDTJ. An abstract in Japanese was presented at the 16th ASINDTJ annual meeting (26 November 2016, Yokohama, Japan), and a short review in Japanese was published in Nihonijishinpo (29 April 2017). The authors thank all the ASINDTJ members who cooperated with this study, especially Dr Hideichi Makino for valuable discussion.
Notes
J Diabetes Investig2018;9:903–907 [Google Scholar]
References
1. Pickup JC. Insulin‐pump therapy for type 1 diabetes mellitus. N Engl J Med2012; 26: 1616–1624. [PubMed] [Google Scholar]
2. Torlone E, Fanelli C, Rambotti AM, et alPharmacokinetics, pharmacodynamics, and glucose counterregulation following subcutaneous injection of the monomeric insulin analogue (Lys(B28), Pro(B29)) in IDDM. Diabetologia1994; 37: 713–720. [PubMed] [Google Scholar]
3. Ross PL, Milburn J, Reith DM, et alClinical review: Insulin pump‐associated adverse events in adults and children. Acta Diabetol2015; 52: 1017–1024. [PubMed] [Google Scholar]
4. Ross P, Gray AR, Milburn J, et alInsulin pump‐associated adverse events are common, but not associated with glycemic control, socio‐economic status, or pump/infusion set type. Acta Diabetol2016; 53: 991–998. [PubMed] [Google Scholar]
5. Attia N, Jones TW, Holcombe J, et alComparison of human regular and lispro insulins after interruption of continuous subcutaneous insulin infusion and in the treatment of acutely decompensated IDDM. Diabetes Care1998; 21: 817–821. [PubMed] [Google Scholar]
6. Cope JU, Morrison AE, Samuels‐Reid J. Adolescent use of insulin and patient‐controlled analgesia pump technology: A 10‐year Food and Drug Administration retrospective study of adverse events. Pediatrics2008; 121: e1133–e1138. [PubMed] [Google Scholar]
7. Kanda Y. Investigation of the freely available easy‐to‐use software ‘EZR’ for medical statistics. Bone Marrow Transplant2013; 48: 452–458. [PMC free article] [PubMed] [Google Scholar]
8. Guilhem I, Balkau B, Lecordier F, et alInsulin pump failures are still frequent: A prospective study over 6 years from 2001 to 2007. Diabetologia2009; 52: 2662–2664. [PubMed] [Google Scholar]
9. Pickup JC, Yemane N, Brackenridge A, et alNonmetabolic complications of continuous subcutaneous insulin infusion: A patient survey. Diabetes Technol Ther.2014; 16: 145–149. [PMC free article] [PubMed] [Google Scholar]
10. Wheeler BJ, Heels K, Donaghue KC, et alInsulin pump‐associated adverse events in children and adolescents–a prospective study. Diabetes Technol Ther.2014; 16: 558–562. [PubMed] [Google Scholar]
11. Guenego A, Bouzillé G, Breitel S, et alInsulin Pump Failures: Has There Been an Improvement? Update of a Prospective Observational Study. Diabetes Technol Ther.2016; 18: 820–824. [PubMed] [Google Scholar]
12. Rabbone I, Minuto N, Bonfanti R, et alInsulin pump failures in Italian children with Type 1 diabetes: Retrospective 1‐year cohort study. Diabet Med2017; 34: 621–624. [PubMed] [Google Scholar]
Articles from Journal of Diabetes Investigation are provided here courtesy of Wiley-Blackwell
The MiniMed™ 640G insulin pump with SmartGuard™ closely mimics the way a healthy pancreas delivers basal insulin to your body. As part of the MiniMed™ system it can guide you towards better control1 by providing advanced protection from hypo events2. All of this with personalised settings to fit your daily life and therapy needs.
LEARN MOREContinuous Glucose Monitoring (CGM) technology tracks glucose levels in your body 24 hours a day. CGM uses a sensor that connects to your MiniMed™ insulin pump giving you early warnings of highs and lows. CGM is clinically proven to lower HbA1c levels1 and reduce the time of hypos3, and aims to improve quality of life.
LEARN MOREThe Ascensia Diabetes Care Contour® NEXT LINK 2.4 is the only blood glucose meter with proven accuracy that sends glucose readings wirelessly to your MiniMed™ 640G insulin pump saving you time and effort. It also allows you to discretely send a bolus directly to your pump and to conveniently download blood glucose and insulin information to CareLink™ Personal, our cloud based therapy management software.
LEARN MOREOur infusion sets with the unique MiniMed™ connection are designed to provide a more secure insulin delivery from your insulin pump to your body. We offer a wide portfolio of different infusion sets to meet your lifestyle needs and body type. Feel comfortable and safe anytime!
LEARN MORECareLink™ Personal Software is a convenient online tool that accurately tracks your insulin usage and glucose levels, allowing you to compare any changes in glucose levels with your daily activities, such as meals and exercise routines. Clear charts and graphs show you these changes and help you and your diabetes healthcare team discover patterns and adjust your therapy for better control.
LEARN MOREWORKING HARD TO KEEP GLUCOSE LEVELS UNDER CONTROL?
Since many factors can affect your glucose levels, such as diet, exercise and stress levels, you may need a system that can give you better control: an intelligent system. The MiniMed™ 640G insulin pump features innovative technology to more closely mimic the way a healthy pancreas delivers basal insulin to the body in order to help you to achieve better control.
Medtronic Minimed 620g
MiniMedTM 640G system also has some other integrated intelligent features for increased protection, such as:
Active insulin tracking
so the insulin still working in your body is displayed on the pump screen and taken into account in the Bolus Wizard™ recommendation
Bolus progress bar
allowing you to see the amount of bolus insulin being delivered; and Stop Bolus, to easily stop a bolus at any time
Predictive battery end of life
so you know when your battery will run out before it happens
PERSONALISED CONVENIENCE - TAILORED TO YOUR NEEDS
Regularly testing your glucose levels. Calculating your insulin dose. Logging your results. There is a lot to think about to maintain daily glucose control. The MiniMed™ 640G insulin pump could make your day easier. It is designed to fit your daily life and take out some of the complexity:
- Personalised basal and bolus settings to fit your needs throughout the day and night
- Programmable friendly reminders so you don’t forget any important activities
- Adjustable audio options so you can change alert notifications depending on your environment e.g., when you are asleep, at the cinema or in a noisy restaurant
- Need help calculating your insulin doses? Bolus Wizard™ can calculate and recommend insulin doses****, displaying the whole calculation on one screen, so you can easily understand it and manage your insulin conveniently and more precisely
- Wireless blood glucose meter (CONTOUR® NEXT LINK 2.4 from Ascensia Diabetes Care) that sends glucose readings directly to your pump and can deliver a bolus remotely
- With our MiniMed™ accessories you can wear your pump the way you want. Discover the collection here
With the design of your MiniMed™ 640G system, you now can:
- Enjoy water activities with our waterproof design+
- Experience easy readability day & night with the full colour and light adjusting display
- Benefit from the intuitive, easy navigation with less button pushing
- Designed for both right and left handed users
- Compared to multiple daily injections, according to the STAR 3 clinical study: Bergenstal RM, Tamborlane WV, Ahmann A, et al. Effectiveness of sensor-augmented insulin-pump therapy in type 1 diabetes. N Engl J Med. 2010;363:311–320
- Garg S, Brazg RL, Bailey TS , et al. Reduction in duration of hypoglycemia by automatic suspension of insulin delivery: the in-clinic ASPIRE study. Diabetes Tech Ther. 2012;14(3):205-209.
- Section 8 clinical study. Data on File. Bayer Healthcare, LLC.
* Waterproof at time of manufacture up to 3.6 meters for up to 24 hours at a time. See MiniMed™ 640G User Guide for a complete description of the waterproof capabilities and proper use instructions
Minimed 620g Insulin Pump
*** The dynamic resume feature is based on certain preset criteria: sensor glucose must be 1.1 mmol/l above the preset low limit and predicted to be 2.2mmol/l above within 30 minutes AND insulin must have been suspended for at least 30 minutes.